FREE SHIPPING ON ORDERS OVER €49.90

Multicare In Glucose Electrode Blood Glucose Test Strips 50 Pieces Multicare

€66,00
Tax included, shipping and discounts calculated at checkout.
Multicare  |  SKU: 937388460  |  Barcode: 8032937230344



Shipping Costs and Times

  • Order fulfilled in 24h
  • Italy, including islands, Free from €49.90
  • Orders under €49.90: €4.99
  • Europe starting from €11

Read the full policy

Payments

Payment information is processed securely. We do not store credit card information or have access to your credit card information.

Returns and Refunds

You can return and receive a refund for the item within 30 days. See full policy here.

Assistenza Clienti BFarma +39 347 093 4913

Do you need help?

Our customer service is here to help you!

  • Contact us by phone, email or WhatsApp, from Monday to Friday, from 9:00 to 20:00 .
  • For common questions, such as tracking your order or checking its fulfillment status, you can count on our artificial intelligence , available 24/7 .

We are always at your side to offer you fast and effective support!

Description

MULTICARE GLUCOSE MEASUREMENT STRIPS ELECTRODE 50 PIECES MULTICARE



multiCare-in
Glucose strips Page

Description
Quantitative whole blood glucose test. For self-monitoring.
SAMPLE: capillary blood. Minimum volume: 0.5 mcL.
HEMATOCRIT RANGE: 30-55%
MEASUREMENT RANGE: 10-600 mg/dL (0.6-33.3 mmol/L)
SENSITIVITY: minimum measurable value: 10 mg/dl
PERFORMANCE FEATURES
REPEATABILITY EVALUATION
The multiCare-in glucose strips showed good performance.
Based on the results, the CV is:
- for glucose concentrations < 100 mg/dL, from 2.4% to 4.3%
- for glucose concentrations ~ 100 mg/dL, 2.0% to 2.8%
EVALUATION OF REPRODUCIBILITY (INTERMEDIATE PRECISION)
The multiCare-in glucose strips showed good performance in all ranges.
Three lots were evaluated, below are the CVs obtained:
Lot. 1: CV between 2.53% and 4.72%
Lot. 2: CV between 2.66% and 4.29%
Lot. 3: CV between 3.03% and 4.29%
1. ACCURACY
Comparison of the multiCare-in blood glucose monitoring system with the reference method using simultaneously collected capillary whole blood samples from 100 subjects.
1.1 Acceptability criteria
Minimum acceptable requirements for accuracy according to ISO 15197:2015.
- 95% of individual blood glucose results must fall within ± 15 mg/dL (0.83 mmoi/L) for glucose concentrations < 100 mg/dL (5.55 mmol/L).
- 95% of individual blood glucose results must fall within ± 15% for glucose concentrations ≥ 100 mg/dL (5.55 mmol/L)
1.2 Results
System Accuracy Results for Glucose Concentrations < 100 mg/dL (5.55 mmol/L)
within ± 5 mg/dl (0.28 mmol/L) within ± 10 mg/dl (0.56 mmol/L) within ± 15 mg/dl (0.83 mmol/L)
80/150 (53.3%) 125/150 (86.6%) 145/150 (96.6%)

System Accuracy Results for Glucose Concentrations ≥ 100 mg/dL (5.55 mmol/L)
within ± 5% within ± 10% within ± 15%
274/450 (60.8%) 411/450 (91.3%) 445/450 (98.8%)

System accuracy results for glucose concentrations between 28 mg/dL (1.5 mmol/L) and 580 mg/dL (32.2 mmol/L)
within ± 15 mg/dL (0.83 mmol/L) or within ± 15% = 590/600 (98.3%) READING TIME: 5 seconds
TEMPERATURE OF USE OF THE STRIPS: 10-40°C

Components
Glucose Oxidase: 21 mcg; Mediator (Hexaammineruthenium chloride): 139 mcg; Buffer5.7 mcg; Stabilizer: 86 mcg.

Warnings
To be used with the instrument.

Conservation
Store at a temperature between 4° and 30°C.
EXPIRY AFTER OPENING: 90 days
EXPIRY: 24 months

Format
Contains: strips; 1 leaflet; 1 data chip.
Pack of 50 strips

Code 3830280

Product Details

  • Format
    Box
  • Shelf life after opening
    90 Days
  • Legislative class
    In vitro Diagnostic Medical Device Directive 98/79/EEC; Commonly Used Medical Device
  • Deductible Product
    Yes